Recently we were asked to supply a prototype of the Auto-Ranging version of our FID Controller. The purpose was to add more analytical channels to a GC from one of the world’s top GC manufacturers.
To speed up the evaluation we had developed a control driver so that the data could be taken digitally in the exact same way as the original built-in GC FID detector.
The results confirmed the qualities of the FIDCU-2DR. Under the exact same conditions, both running at 50Hz sampling frequency, using the same sample and the same physical FID hardware, the Applied Instruments FID interface showed a S/N ratio that was at least 2-3 times better than the original manufacturer’s FID electronics! The screen shots below show the results. The blue trace is the Applied Instruments FID interface signal where the green trace shows the original equipment FID electronics.
To make these results even more impressive; the Applied Instruments FIDCU-2DR was lying on top of the GC close to the heated zones whereas the original FID electronics were nicely sheltered inside the GC cabinet. |